CBC with Auto Differential
Additional Codes
LAB1748
Performing Laboratory
MBMC Main Laboratory
Methodology
Quantitative Automated Hematology Analyzer
Testing Algorithm
If platelet count is < 50 x103 /uL, the Immature Platelet Fraction (IPF) will reflex with an additional charge.
Specimen Requirements
Specimen Type: Whole blood
Container/Tube: Lavender top (EDTA)
Specimen Volume: Full tube
Specimen Minimum Volume: 1 mL
Collection Instructions: 500 microliters of blood in a lavender EDTA Microtainer is also acceptable.
Specimen Transport Temperature
Ambient
Day(s) Test Set Up
Monday through Sunday
Turnaround Time:
STAT: 1 hour after receipt in laboratory
Routine: within 4 hours
Reference Values
See Reference Ranges in Special Instructions
|
Hematology Criticals |
Rule |
|
|
Platelets |
||
|
Inpatient |
Less than 20 or Greater than 1000; |
|
|
Inpatient |
Less than 25 or Greater than 1000; |
|
|
ED |
Less than 50 or Greater than 1000; |
|
|
Outpatient |
Less than 50 or Greater than 1000; |
|
|
MBMC Labor and Delivery Adult ≥18 years |
Less than 50 or Greater than 1000; |
|
|
MBMC Cancer Center Adult ≥18 years |
Less than 20 or Greater than 1000; |
|
|
Hemoglobin |
||
|
Outpatient |
Less than 7.0; |
|
|
Outpatient |
Less than 6.0; |
|
|
Outpatient |
Less than 6.5 or Greater than or equal to 20; |
|
|
Inpatient/ED |
Less than 7.0; |
|
|
Inpatient/ED |
Less than 6.0; |
|
|
Inpatient/ED |
Less than 6.5 or Greater than or equal to 20; |
|
|
WBC |
||
|
Outpatient |
Less than 1.0 or Greater than 50.0; |
|
|
Inpatient/ED |
Less than 1.0 or Greater than 50.0; |
|
|
Inpatient and Outpatient/ED |
Less than 2.0 or Greater than 50.0; |
|
|
Differential |
|
|
|
Outpatient |
Blasts >0% |
|
|
Inpatient/ED |
Blasts >0% |
|
|
Inpatient/ED |
Blasts >0% |
|
|
Inpatient/ED |
NEUTAB <0.50; |
|
|
Outpatient |
NEUTAB <0.50; |
|
|
Inpatients/ED |
ANC <0.50; |
|
|
Outpatients |
ANC <0.50; |
|
Test Classification and CPT Coding
Test Classification:This test has been cleared or approved by the U.S. Food and Drug Administration and is used per manufacturer’s instructions.
CPT Code:
85025
85055 - Immature platelet function, if indicated
Stability
24 hours
Profile Information
Includes WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, platelet count, NRBC, MPV, and 5-part automated differential (% absolute neutrophils, % lymphs, % monos, % eos, and % baso).
A manual, 100-cell differential count is available ONLY when generated by reflex criteria established by MBMC Laboratory. Notify Laboratory if ADD ON manual differential is ordered by physician (EPIC CODE: LAB6668) within 24 hours of collection. All alert values will be called and documented.If an abnormal value or flag is triggered on the initial patient specimen, a manual smear will always be reviewed. Any subsequent specimen within 72 hours will be verified by the “Consistent Rule” unless there is a change in the values and flags in which case a new smear review will be conducted.