Pancreatobiliary FISH
Performing Laboratory
ARUP Laboratory
Specimen Requirements
Collect: Bile or pancreatic duct brushings, biliary stent, or fine needle aspirates of the pancreas in UroVysion FISH Collection Kit (ARUP Supply #41440) available online through eSupply using ARUP Connector contact Client Services at (800) 522-2787. For specific instructions refer to Specimen Collection & Handling.
Specimen Preparation: Place specimen in Cytolyt or PreservCyt fixative vial. If the specimen is a brushing, submit the brush in the fixative. Specimen should be placed in fixative vial immediately after collection.
NOTE: Specimen source is required.
Specimen Transport Temperature
Storage/Transport Temperature: Ambient or Refrigerated.
Unacceptable Conditions: Frozen specimens. Unfixed specimens not collected in Cytolyt or PreservCyt fixative. Specimens submitted in expired collection vials.
Stability: Ambient: 1 week from collection; Refrigerated: 1 week from collection; Frozen: Unacceptable
Day(s) Test Set Up
Perfomed: Mon-Fri
TAT: Reported 4-12 days
Reference Values
Positive | ≥5 cells with gains of two or more chromosomes are present |
Equivocal Tetrasomy | Cells present with tetrasomic or near tetrasomic signal patterns |
Equivocal Trisomy | ≥10 cells with gains of a single chromosome, (Trisomy 7 or Trisomy 3) |
Negative | < 5 abnormal cells are present |
Test Classification and CPT Coding
88366